LIBS calibration

To calibrate our LIBS spectrometer measurements to correspond to real peak wavenumbers and intensities we can use the following methods:

Wavelength calibration

To calibrate a spectrometer resulting wavelengths we use the following method. We plot a theoretically calculated(see: Artificial references) element LIBS spectrum and a measured element spectrum in the same figure. We adjust the fitting function, a second degree polynomial, parameters A, B and C until the plots match each other and save the found out three number parameters.

For a calibration target we use peaks of elements O and N which are always present in the air.

figure_calibration figure_calibration2 figure_calibration3

Note that in our example results we often use combined measurements from two different spectrometers and so this calibration is done separately to each spectrometer. For some instruments a 2-degree polynomial fitting might not make a good match. Some other function fit or lookup-table might be necessary.

Intensity calibration

The light intensity for each wavelength of light is different for differing optical setups and spectrometers. Therefore a calibration curve to bring your device signal levels to a "standard" is a good idea. Spectrometer sensitivity curves are a good starting point but for accurate results you might need to tweak it.

Sensitivity curves for our spectrometers(Avantes and BWSpec)

Manufacturer provided figure_

Experimental (Calculated to get best fit for our models) figure_

Element concentration calibration

Methods for this are presented in the section for quantitative elemental analysis.

CC BY-SA 4.0 Ilkka Laine. Last modified: January 20, 2026. Please contact me by email for any questions, suggestions or improvements.